Port Saver™ Instrument Protection DC Block


Port Saver™ Instrument Protection DC Block
P2131A – Standard
Characteristic | Rating |
---|---|
Absolute Maximum Input Voltage | +/-75VDC |
Typical 3dB Frequency Range | 200Hz – 1GHz |
Pre-change, Dis-change Time Constant | 10s |
Insertion Loss (Typical) | 0.8dB |
Probe Factor | 1.1 |
Return Loss | <1.3 |
Impedance | 50 Ohms |
Instantaneous Peak Inrush Current | 15 Amps |
Inrush Clamp Voltage | 7.5V < 1us |
Connector Type | SMA (Male) – SMA (Female) |
Working Temperature | -40 -+75 C |
P2132A
Characteristic | Rating |
---|---|
Absolute Maximum Input Voltage | +/-6VDC |
Typical 3dB Frequency Range | 1.3Hz – 450MHz |
Pre-change, Dis-change Time Constant | 60s |
Insertion Loss (Typical) | 12.1dB |
Probe Factor | 4.1 |
Return Loss | <1.3 |
Impedance | 50 Ohms |
Instantaneous Peak Inrush Current | 1 Amp |
Inrush Clamp Voltage | 250mV < 1us |
Connector Type | SMA (Male) – SMA (Female) |
Working Temperature | -40 -+75 C |
Note: Blue bold text indicates a Characteristic that can be customized
P2131A Port Saver | DC Blocker 500Hz – 800MHz (Works with all 50 ohm equipment, 75VDC maximum differential voltage) | $495.00 Add to Cart |
P2132A Port Saver | DC Blocker 2Hz – 50MHz (Works with all 50 ohm equipment, 6VDC maximum differential voltage) | $495.00 Add to Cart |
The Port Saver™ family of DC blocks incorporates new features designed to protect your most valuable instruments while also greatly improving performance. Port Saver includes inrush currently limiting, which together with the transient protection, limits the voltage that appears at the instrument when a DC voltage is applied. Pre-charge allows the block to charge even without a connection to the instrument, eliminating the inrush. Self-discharge allows the DC block to discharge even if it is removed from both the source and the instrument, eliminating potential damage from energy stored in the block.
Typical RF blocks are based on ceramic capacitors, which exhibit AC and DC bias effects, making it very difficult to calibrate, as well as ferro-resonant noise. Port Saver blocks do not exhibit DC or AC bias effects, resulting in greatly improved calibration and measurement accuracy and do not exhibit ferro-resonant noise, making them quieter.
Port Saver DC blocks can easily be customized for different voltages and bandwidth. Please inquire with Picotest at info@picotest.com

P2131A and P2132A Insertion Loss – Uncalibrated
Port Saver™ Instrument Protection DC Block
Description | Type | Size |
---|---|---|
User Manual for Picotest PDN Probes and Port Saver DC Block | 3.36 MB |